Figure 6: lel detector wiring • co or h, Eclipse detector head operator’s manual 9, Black red – RKI Instruments Eclipse User Manual
Page 9
Advertising

Eclipse Detector Head Operator’s Manual
9
Figure 6: LEL Detector Wiring
•
CO or H
2
S gas detector.
Red wire to terminal labeled TOXIC RED, black wire to terminal labelled TOXIC
BLK.
Figure 7: H2S/CO Detector Wiring
•
Oxygen Detector.
White wire to terminal labelled OXY WHT, green wire to terminal labelled OXY
GRN.
Figure 8: Oxygen Detector Wiring
GRN
WHT
White
Black
Red
Green
RED
GRN
WHT
ECLIPSE
BLK
XXX
BLK
+
F B
-
24 VDC
IN
X X X
Cl2:
Yellow
SO2: Orange
CO:
Red
H2S: Red
WHT
GRN
Black
Red
RED
WHT
GRN
XXX
BLK
+
2 4 V D C
I N
F B
-
ECLIPSE
X X X
Cl2: Yellow
SO2: Orange
CO: Red
H2S: Red
BLK
WHT
Green
BLK
XXX
GRN
GRN
-
WHT
RED
F B
24 VDC
IN
+
ECLIPSE
X X X
Cl2: Yellow
SO2:Orange
CO: Red
H2S: Red
BLK
White