Model GD-K8A Diffusion Detector Head • 13

7.

Remove the milliampmeter test clips from the output check pins.

8.

Close the housing door.

9.

Allow time for the gas reading to return to normal, then return the controller to
normal operation.

NOTE:

If the gas reading does not return to normal before returning the controller to
normal operation, unwanted alarms may occur.

10. Store the components of the calibration kit in a safe and convenient place.

11. Verify that the controller display reading decreases and stabilizes at 0.