RKI Instruments GX-2001 User Manual

Page 7

Advertising
background image

Data Logging Software User’s Guide

Overview of the Data Logging Software • 7

the Model GX-2001 going into and out of alarm, but you may not have five
scheduled log times.

The information on the previous page can also be displayed in summary
form by clicking the Summary button.

Alarm Trend Data

In addition to the interval trend data the GX-2001 also saves a file which
describes the most recent event during a particular session.If an event
occurs, then an alarm trend file is saved separately from the data file, which
shows the reading 15 minutes before and 15 minutes after the event, with
the log time every five seconds.

When an event occurs and an alarm trend data file is saved in the GX-
2001, this file remains in the unit’s memory until another event occurs.
When another event occurs, the alarm trend data file is overwritten with the
new event’s information.

Alarm trend data can always be displayed in either table and graph format.

Figure 2. Data View (Interval) Summary Window

Gas concentrations

Warning levels

Time log

(date and time)

Advertising