OPERATION

All display lamps are turned on and off with a single switch. Each lamp is

supplied with an eight-foot cord and plus brackets for hanging the lamp.

REPLACING A FIL

TER

Shortwave filters are subject to deterioration due to solarization, and need

replacement periodically. Depending on the application, a lifetime of 500 to

1000 or more hours of operation can be expected. A

display lamp is normally

in a fixed position. Not all fluorescent minerals have the same brightness

response. The brighter specimens will often respond for over 1000 hours if

within a reasonably close range, but if you are showing specimens of lower

level of fluorescent response, you may have to bring the lamp closer or

replace filters earlier than normal.

T o change the filter

, remove the two screws on the filter frame and remove

end bracket. Replace the filter/frame assembly and screws.

REPLACING THE TUBES

T o replace the tubes in the lamp

s, remove the two screws on each side of the

unit and remove the filter cover. Carefully twist tubes out using care when

removing. Note that the reflector is not screwed into the housing. If lamp

does not light after replacing the bulbs, replace the st

arter. For any other

questions, contact UVP’s offices.

MOUNTING INSTRUCTIONS

All lamps come with wall mounting brackets attached to the housing. Simply

bolt or screw the brackets to the desired location. The lamp can be mounted

horizontally or vertically

.

REPLACEMENT PARTS

Filter/Frame

Model

Assembly

Tubes

UVL-225D

38-0207-02

34-0060-01 (365nm Longwave)

UVS-225D

38-0207-01

34-0073-01 (254nm Shortwave)

UVLS-225D

38-0207-01

34-0073-01 and 34-0060-01

UVM-225D

38-0207-01

34-0072-01 (302nm Midrange)

REPAIRS

Contact UVP’

s offices with any questions.

T o return a unit to the factory for

repair

, call UVP’s customer service dep

artment for a Return Authorization

Goods (RGA) number prior to sending the unit.

ACCESSORIES

Blak-Ray®

Safety Goggles and Contrast Control Spectacles ... Eyewear

is recommended when using any shortwave UV product. Special formula

lenses reduces “blue haze” interference while protecting eyes form harmful

bands of UV . UVC-503 Goggles provide maximum protection from extended

or high intensity UV light sources while the UVC-303 Spectacles are used for

sporadic lower intensity UV light sources and can be worn comfortably over

prescription glasses.

Ultraviolet Intensity Meters ... For widest energy range measurements,

highest accuracy , and interchangeable sensors (ordered sep

arately) for mea-

surements at 365nm, 300nm and 254nm, the new UVX Digital Radiometer

can be used. Units are hand-held, battery operated, and have compact sen-

sors with 3' electrically shielded cord. Also available are Blak-Ray Meters in

models J-221, which measures longwave (365nm), or J-225 shortwave (254nm)

ultraviolet. Compact (fits in the hand). Removable sensors call for ease of

operation. Highly accurate. For measuring ultraviolet intensity from ultraviolet

sources.