5. Maintenance

Daily maintenance/Weekly maintenance

Do not disjoint the unit

Never modify the unit. There are some high-

voltage areas inside the unit, which could cause

electric shock. Request Yamato Scientific for

internal check, adjustment or repair.

Maintenance

Daily Maintenance

:

Turn power switch off and disconnect main power cord before attempting any maintenance.

Once the water bath has cooled to room temperature you may empty the water.

Wipe off any excess dirt on the interior and exterior with a damp cloth. Do not use benzene,

thinner or cleanser.

Weekly Maintenance

:

The pipe heater should be cleaned at least twice a month. Accumulation of scale build up

results in poor heat transfer and high sheath temperatures.

When the unit is not in use for a long period of time.

Be sure to disconnect main power and drain the water tank if the unit will not be used for a long

period of time.

Troubleshooting Guide

Situation

Make Sure

The power lamp on the panel is not

lit when the power switch is on.

● Check main power cord is connected firmly.

● Check main input voltage.