Operation Method
Temperature Output Terminal
Specification
Temperature Output
(ANALOG)
• The voltage (DC) corresponding to the measured temperature is output.
• Output temperature range: CLS301 -15 to 35℃, CLS400/600 -20 to 35℃
• Output voltage: 4 to 20mA DC
• Load:
600Ω or bellow
• Resolution: ±1℃
• Connection: M4 screw terminal block
CLS301
Temperature Output
-15
-2.5
10
22.5
35
4
8
12
16
20
Output Voltage (mA)
Temp.
CLS400/600
Temperature Output
-20
-6.25
7.5
21.25
35
4
8
12
16
20
Output Voltage (mA)
Temp.