Operation method, Program operation, Program creation – Yamato Scientific DKN 912 Constant Temperature Drying Oven User Manual

Page 29

Advertising
background image

25

Operation Method

Program Operation

Program creation

The program pattern below is explained as an example.

1. Program pattern example

Temp.

200℃

100℃

0℃

Step

1 2 3 4 5 6 7

8 9

Temp.(℃)

100 100 150

150 200 200

50

50

25

Time(min.)

25 30 15 90 25 180

150

← Repeat function →

The number of steps is not counted.

180

100

POWER

1. Turn on the power

• Turn on the power switch of the unit.
• The display on the controller lights on.
• The initial screen is displayed for about four seconds, then the

measurement temperature (temperature in furnace) is displayed.

™ The initial screen displays the software version information, sensor

used and setting temperature of overheating prevention device.

2. Select program mode/program pattern

① Press the PROGRAM key once.

The measurement temperature display screen displays the
previous program mode.
Press the PROGRAM key again to display the next program
mode.

② Select the mode and press the ENTER key.
• When PrG1 is selected, the measurement temperature display

screen displays "End".

• When PrG2 is selected, the measurement temperature display

screen displays the program pattern "PAt1". For the pattern of
PrG2, select "1" or "2" using the "▲▼". Press the ENTER key
again. The measurement temperature display screen displays
"End".

• When PrG3 is selected, the measurement temperature display

screen displays "PAt1". For the pattern of PrG3, select "1", "2"
or "3" using the "▲▼". Press the ENTER key again. The
measurement temperature display screen displays "End".

™ Any of PrG1. PrG2 or PrG3 can be selectable in the program

example above, where nine steps maximum are used.

Advertising