Operation method, Operation mode and function list – Yamato Scientific IQ820 Low Temperature Incubators User Manual

Page 17

Advertising
background image

14

Operation Method

Operation Mode and Function List

The operation function of this unit is as follows;

No.

Name

Description

Page

Auto overheating
prevention function

This function is set to be automatically activated
(auto reset) when the temperature exceeds the
setting temperature by 6℃.

1.

Overheating
prevention
function

Overheating
prevention device

Though the device shares power source, display,

and key input with the controller, it has
independent temperature measurement circuit,
CPU, sensor and output circuit. Overheating
prevention temperature can be set using the
operation panel.
The unit stops operation when the device is

activated. The unit starts operation again when
the POWER switch is pressed again (manual
reset).

16

2.

Calibration offset function

This calibration offset function is for calibrating

the difference occurred between the required in-
furnace temperature and control temperature
(sensor temperature) of the controller. This unit
can be calibrated toward either plus side or minus
side of the whole temperature range.

30

3.

Overheating prevention
temperature calibration
function

The temperature of overheating prevention device
is automatically corrected when the temperature
of controller is collected.

-

4.

Recovery at power failure

The unit starts operation with the same condition
as just before power failure if it occurs during
operation.

-

5.

Setting value locking

This function locks the established operation
status.
It can be set and cancelled with the SUBMENU
key.

30

Advertising