2 Installation

818 IC Pump / 8.818.1023 Instructions for Use

10

6.2744.010 PEEK compression fittings can be used. The flow direction

is arbitrary. The pulsation dampener is positioned in the interior of the

820 IC Separation Center on the base below the injection valve (see

819/820 Instructions for Use).

The pulsation dampener is filled with isopropanol when new. Rinse

your IC system carefully after the first installation of a new pulsation

dampener.


2.3.3

Filter unit PEEK

The 6.2821.120 Filter unit PEEK 24 supplied (see Fig. 4) serves to

avoid contamination of the piston seals by abrasive particles and can

be used in the pressure range 0

30 MPa (0

300 bar). The filter unit

consists of the housing 21, the filter 22 and the connector 23 to be

screwed into the housing 21. For the connection of capillaries 19, PEEK

compression fittings 20 (6.2744.010) must be used.
New filter 22 are available as an option with the ordering number

6.2821.110 (10 pieces).

For the connection of the filter unit, please note the flow direction

arrow printed on the housing.
The filter unit is filled with isopropanol when new. Rinse your IC system

carefully after the first installation of a new filter unit.

Fig. 4:

Filter unit PEEK (6.2821.120)

19 PEEK capillary (6.1831.010)

20 Compression fitting (6.2744.010)

21 Filter housing

22 Filter 6.2821.130

23 Filter connector

20

19 20

23

22

21 19