2 suitability test for tubing materials, 3 tubing dimensions, 4 tubing lifetime – Metrohm 772 Pump Unit User Manual

Page 25: Suitability test for tubing materials, Tubing dimensions, Tubing lifetime

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Operation and maintenance

772 Pump Unit

■■■■■■■■

17

5.2.2

Suitability test for tubing materials

Place a piece of the tubing in question for 48 hours in a closed container
filled with the liquid in question. Then examine the piece of tubing for
signs of swelling, embrittlement or other damage.

5.2.3

Tubing dimensions

Note

Use tubing with a wall thickness of 1.6 mm only.

The largest usable inner diameter is 6.4 mm.

5.2.4

Tubing lifetime

The pump tubing has a limited lifetime and therefore has to be replaced
regularly. The following are examples of factors that can influence the life-
time of tubing in peristaltic pumps:

■

Chemical resistance of the pump tubing

■

Pump rate (rotational speed)

■

Properties of the pumped liquid

■

Outlet pressure

■

Ambient temperature

■

Shearing force of the rollers

■

Twisted or kinked tubing

■

Tubing too long

■

Wall thickness of 1.6 mm not complied with

Advertising