3 equipping the supply bottle, Equipping the supply bottle – Metrohm MiPT User Manual

Page 15

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

IC-Ausrüstung: MiPT

■■■■■■■■

9

3.3

Equipping the supply bottle

Required accessories

■

Bottle (6.1608.070) filled with ultrapure water

■

Eluent bottle cap (6.1602.160)

■

Adsorber tube (6.1619.000), included in the accessories for the dosing
unit (6.3032.120)

■

Adapter for adsorber tube (6.1624.000)

■

FEP aspiration tubing (6.1819.110)

■

M8 stopper (6.1446.080), included in the accessories for the eluent
bottle cap (6.1602.160)

1 Mounting the aspiration tubing

■

Insert the aspiration tubing into the M6 opening of the eluent
bottle cap.

■

Use the capillary cutter to cut the aspiration tubing to such a
length that it touches the bottom of the bottle.

2 Inserting the stopper

■

Tighten the M8 stopper in the M8 opening of the eluent bottle
cap.

3 Mounting the adsorber tube

■

Fill the adsorber tube with some cotton and adsorber material.

■

Place the adsorber tube on the adapter.

■

Insert the adapter into the SGJ opening of the eluent bottle cap.