6.12 793 Sample Prep Module
IC Net 2.2
105
First column
Time at which program instruction is ap-
plied.
Entry range:
0.0 ... 999.9 min
If no time is entered, the program instruc-
tion is applied together with the last instruc-
tion with time entry.
Second column
Program instruction (see below).
Third column
Parameter for program instruction (see be-
low).
ENABLED
Enable program start (a disabled program
is not started).
<Add>
Add new program instruction.
<Delete>
Delete selected program instruction.
<Verify>
Test the time program (error messages are
displayed if program is wrong).
List of program instructions
The following program instructions can be added to the time pro-
gram on the
Program
page:
Instruction Parameter
entry Meaning
Program RESET,
END
Set program flag. The
END
flag
can be used to end a program,
especially if the program time
should be longer than the chro-
matogram
duration
. Additional
steps after this flag are not al-
lowed. The
RESET
flag is used to
reset the parameters to the sys-
tem startup values.
Pump on,
off
Switch on or off the pump drive.
Step
Rotate sample preparation mod-
ule to the next position.
Links
The
Links
tab of the
793 Sample Prep Module
is used for Event line
selection.