Figure 4

Interconnecting Wiring

17 18 19 20

21 22

23

24

TB3

LINE

115/220

VAC

GREEN

WHITE

BLACK

TB1

6 7

8

0-10V

OUTPUT

TB2

9 10 11 12 13 14 15 16

(1) RELAY CONTACTS AT TB3

17 & 18 AND TB3-20 &21

ARE CLOSED IN THE RECIR-

CULATE MODE. RELAY CON-

TACTS AT TB3-18&19, AND

TB3-21&22 ARE CLOSED IN

THE OVERBOARD MODE.

(2) INTERCONNECT CABLE IS A

8 WIRE, 20 AWG., SHIELDED

CABLE. BLUE WIRE NOT USED.

YELLOW

ORANGE

WHITE

GREEN

BLACK

RED

SHIELD

REMOTE

SENSOR MODULE

EARTH

GROUND

E9

+_

NOTES:

-

+

4

3

1

2

3-32VDC

120/240VAC

TO

CONTROL CIRCUIT

POTTER & BRUMFIELD

10 AMP

ZERO CROSSING

SOLID STATE RELAY

SSRT-240D10

+

-

VC

VCC

+

-

BROWN

MANHATTAN CABLE #M4665.

(OPTIONAL)

(3) JUMPER WIRE SHOWN FROM TB3

(21) TO (24) USED ONLY WITH

OPTIONAL ZERO CROSSING RELAY.