2.13

Valco MSVA

The MSVA is used with a two position air actuator, which requires
one interval to switch to the inject position and another to switch to
the load position. The steps described below are illustrated in
Figure 3.

1.

Supply the proper voltage (110 VAC in this example) by
connecting the HOT terminal to C of the one of the groups of
terminals to be used, and jumping it to C of the other.

2.

Connect one wire of each solenoid to the NEUT terminal.

3.

Connect the other wire of each solenoid to the NOs which
correspond to the Cs of Step 1, according to the desired
switching sequence.

2.14

Pulsed Operation of Solenoids

The first three procedures demonstrate the simplest uses of the DVSP. This
example will illustrate the use of the pulsed relay in conjunction with two devices
on the same DPDT relay. The steps described below are illustrated in Figure 4.

1.

Connect the two wires of the 12 VDC
solenoid as described in Section 2.12.

2.

Connect one wire of the 110 VAC sole-
noid to NEUT as described in Section
2.11, but connect the other wire to the

second

NO terminal of the

same

relay

to which the 12 VDC solenoid is con-
nected.

3.

Supply 12 VDC by connecting the +12
terminal to one of the NO terminals on
the pulsed relay (Figure 1) and
connecting that NO’s related Common
to the Common of the NO which has
the 12 VDC solenoid connected to it.

4.

Likewise, supply 110 VAC by connecting
the HOT terminal to the remaining NO
terminal on the pulsed relay and
connecting that NO’s related Common
to the Common of the NO which has
the 110 VAC solenoid connected to it.

12 VDC SOLE-

NOID

110 VAC SOLE-

NOID

TERMINALS FOR
EXTRA RELAY
WITH TWO SEC-
OND PULSE

Figure 4: Use of the pulsed relay and

two events on one relay

Figure 3: 110 VAC Valco MSVA