Introduction, Description and operating principle – VICI D-2-IM User Manual

Page 5

Advertising
background image

1

Introduction

Description and Operating Principle

The pulsed discharge detector Model D-2-IM is a universal helium photoionization
detector optimized for trace level work with micropacked and capillary columns.
A schematic representation is shown in Figure 1.

As the name implies, a stable, low power, pulsed DC discharge in helium is utilized
as the ionization source. Elutants from the column, flowing counter to the flow of
helium from the discharge zone, are ionized by photons from the helium discharge
above. Resulting electrons are focused toward the collector electrode by the two
bias electrodes.

The principal mode of ionization is photoionization by radiation arising from the
transition of diatomic helium

He

2

(A

1

+

)

u

to the dissociative 2He(1S

1

) ground state.

This is the well-known Hopfield emission. The photon energy from the He

2

continuum is in the range of 13.5 eV to 17.7 eV.

The PDHID is essentially non-destructive (0.01 - 0.1% ionization) and highly
sensitive. The response to organic compounds is linear over five orders of
magnitude with minimum detectable quantities (MDQs) in the low or sub
picogram range. The response to fixed gases is positive (the standing current
increases), with MDQs in the low ppb range.

Figure 1: Schematic of the D-2-IM detector

BIAS ELECTRODE

COLLECTOR ELECTRODE

BIAS ELECTRODE

CAPILLARY COLUMN

GROUND

QUARTZ
INSULATOR

SAPPHIRE
INSULATORS

VENT

COLUMN INLET

SPRING WASHERS

DISCHARGE ZONE

HELIUM INLET

DISCHARGE ELECTRODE

Advertising