Entering commands – VICI 150 Dynacalibrator User Manual

Page 13

Advertising
background image

11

10. Click “ASCII Setup” and set the parameters as indicated below.

Click “OK”.

11. When the screen returns to Dyna150 HyperTerminal (page 9, bottom

illustration), click “File”, then “Save” to save this Dynacalibrator Model
150 communications setup.

The following text will be displayed in the HyperTerminal window when
communication is first established with the Dynacalibrator Model 150:

Parameters recovered.

Program checksums ok.

Initializing SPI . . . ok

Initializing A/D . . . ok

Initializing IIC . . . DS1624 started

Valco Dynacalibrator online.

A simple processor test can be performed by opening Hyperterminal before
powering up the Dynacalibrator. If the processor has some defects, the
message “Bad checksum in Block xxx” is displayed instead of line 2 above.
If this occurs, please consult the factory.

Entering Commands

Each command must be followed by the <ENTER> key.

Examples:

To set the chamber temperature to 50°C, type “TS=50” and press <ENTER>.
To read the set point, type “TS” and press <ENTER>.
To read the current chamber temperature, type “TR” and press <ENTER>.
To start data logging, type “L+” and press <ENTER>. To stop data
logging, type “L-” and press <ENTER>.

Serial Port Communication

Advertising
This manual is related to the following products: