3

Specifications

Introduction

Operational

Flow controls

Output concentration range

Fractional ppb to hundreds of ppm

Carrier flow rate (nominal)

100 – 400 sccm

Dilution flow rate range(nominal)*

1 -20 SLPM *(depending on model)

Dilution flow accuracy

+/- 1% of Setpoint from 20%-100% of full scale,
+/- .2% of full scale between 2% - 20% of range

Operating temperature range

10° to 50°C (50° - 122°F)

Permeation Chamber

Temperature range*

30° - 110°C *(Must be at least 2°C above ambient)

Temperature setpoint accuracy

+/- 0.05°C (NIST-Traceable) from 30°C -110°C

Temperature setpoint repeatability

+/- 0.01°C at any fixed ambient temperature

Temperature equilibrium time

1.5 hours for highly dynamic changes

Chamber size

Accepts devices up to 23.5 cm X 1.6 cm diameter

Modes

(Auto or Manual) Span Out

Operating Duty Cycle

Continuous

Environmental

Operating noise emission

45 to 50 dBA

Ambient operating temperature

20°C to 35°C

Operating humidity

0-95% relative humidity

Storage temperature

10° C to 40°C

Storage humidity

0 – 50% relative humidity

Power requirements

Voltage, frequency, power

100-240 VAC, 50/60 Hz, 150 Watts maximum