Shutdown Procedure

Dynacalibrators are designed to operate continuously; however, if it is necessary
to turn the unit off, use the following procedures:

1. With the tool provided, rotate the panel lock screw 90° counterclockwise.

2. Gently pull the front panel assembly out.

CAUTION: The permeation chamber cap may be hot. If the
calibrator has been in use, avoid exposure to gas vapors
while opening the chamber cap.

3. Unscrew the chamber cap with the tool provided.

4. Remove and contain permeation devices in an appropriate manner as

described in permeation device literature..

5. Secure the chamber cap with the tool provided,

6. Reinstall the front panel assembly and rotate the panel lock screw

90° clockwise.

7. Place devices in the containers provided and store them in an appropriate

place until further use is required. (A refrigerator or freezer is recommended.)

8. Allow the carrier flow to purge the chamber for about five minutes

9. If the unit will be left unpowered for an extended length of time, disconnect

all the plumbing from the rear panel inlets and outlets and cap them.

When repowering the unit, always perform the start-up procedures starting on
page 14.