LaMotte DC1100 Colorimeter User Manual
Page 7

TROUBLE SHOOTING GUIDE
PROBLEM
CHECK
ACTION
Meter won’t
come on
Battery?
Replace, see pg. 6 of manual.
^
AC Adapter?
See pg. 6 of manual.
^
AC Wall Outlet?
Verify power source.
^
Contact LaMotte for
return authorization.
Return to LaMotte for repair.
^
Comes on, but
readings drift
Are instrument grade
reagents being used?
Obtain instrument grade reagents.
^
Are there suspended solids
in sample?
Filter both blank and sample
(after reaction).
NOTE: Do not use filtered blank for
sample reaction.
^
Do reading drift up scale? Bring samples to room temperature.
^
Readings drift down scale
with cold sample?
Check for sample tube fogging, let
sample equilibrate to room temperature.
Suspect
Calibration
Will meter blank? (zero)
If no, see next page.
^
Are instrument grade
reagents being used?
Obtain instrument grade reagents.
^
How old are the reagents? Contact LaMotte to verify age.
^
Check calibration with
standards?
How old is the standard?
^
Is class “A” volumetric
glassware being used to
make up the standards?
If not, obtain and see Standard
Methods.
7