Lamotte company – LaMotte 1200 COLORIMETER IRON 1200-FE User Manual

Page 4

Advertising
background image

IRON TEST METHOD SPECIFICATIONS

APPLICATION

Drinking , surface, and saline waters; domestic and industrial wastes.

RANGE

0 to 4.0 ppm Iron

METHOD

Ferric iron is reduced to ferrous iron and subsequently forms a colored complex with
phenanthroline for a quantitative measure of total iron.

HANDLING & PRESERVATION

The sample container should be cleaned with acid and rinsed with deionized water.
Addition of acid to adjust the sample to pH 2 - 3 will prevent deposition of iron on the
container walls. Samples should be analyzed as soon as possible after collection since
ferrous iron undergoes oxidation to ferric iron.

INTERFERENCES

Strong oxidizing agents, cyanide, nitrite, phosphates, chromium, and zinc in
concentrations exceeding 10 times that of iron; cobalt and copper in excess of 5 mg/L,
and nickel in excess of 2 mg/L. Bismuth, cadmium, mercury, molybdate, and silver
precipitate phenanthroline.

LaMOTTE COMPANY

Helping People Solve Analytical Challenges

®

PO Box 329 • Chestertown • Maryland • 21620 • USA

800-344-3100 • 410-778-3100 (Outside U.S.A.) • Fax 410-778-6394

Visit us on the web at www.lamotte.com

63681-01

·

4/09

Advertising