LaMotte Tracer SAL Meter User Manual

Page 17

Advertising
background image

TROUBLESHOOTING

Problem

Check

Action

Reading is frozen

HOLD mode

Press MODE/HOLD button
to exit HOLD mode

“BAT” message

Batteries low

Replace batteries

Meter will not
calibrate in
conductivity mode

Trapped air
bubbles

Tap probe or stir sample to
release air bubbles

Dirty probe

Clean conductivity probe

Damaged probe

Replace probe

Contaminated
conductivity
standards

Use fresh standards

Meter will not turn on Batteries low or

dead

Replace batteries

Battery polarity

Replace batteries with
correct polarity

Unit will not respond
to any key presses

Internal fault

Perform hard reboot.
Remove batteries, hold
ON/OFF button down for 5
seconds, replace batteries

"_oL_" message

Sample is out of
range

Dilute sample

17

Advertising
This manual is related to the following products: