TROUBLESHOOTING

Problem

Cause

Action

Power on
but no display

Batteries

Insert batteries

Batteries

Verify correct polarity

Batteries

Replace batteries

Unstable
readings

Electrode

Immerse electrode more deeply
in sample

Electrode

Condition electrode before fi rst
use

Electrode

Remove air bubbles caught
under electrode

Electrode

Clean electrode

Electrode

Replace electrode

Slow response
time or reading
drift

Clogged junction

Soak in 4.07M Potassium
Chloride (KCl) at 60°C for 30
minutes

Strong alkaline
measurement

Soak in 0.1M Hydrochloric acid
(HCl) overnight

Deteriorated gel layer

Replace electrode

Protein coating on
electrode surface

Soak in 1g Pepsin dissolved in 10
mL of 0.1M HCl for 30 minutes
or as needed

Oil, paint, dyes,
suspended solids on
electrode

Rinse electrode alternately with
materials solvent then a buffer
7.00

Dry bulb

Dehydrated membrane,
long term storage
without wetting

Soak electrode tip in wetting cap
fi lled with 1 mL 7.00 buffer for
24-48 hours

Static charge

Wiping electrode

Rinse electrode in 7.0 buffer and
blot. Do not wipe electrode

Same readings
in different
buffers and
samples

Cracked or broken
electrode

Replace electrode. Use bulb
guard. Avoid plunging electrode
into bottom of container and stir
bars.