Hold – LaMotte Tracer ORP Meter User Manual

Page 7

Advertising
background image

ORP Measurement
1. Press the ON/OFF button to turn the meter on. Place the electrode

½” to 1” into the test sample .

2. Stir once and let the reading stabilize for 1 to 3 minutes.

The main display will indicate the ORP in numeric units from 0 to
1000mV. The bar graph will also display the ORP value. As the
ORP increases, the bar graph will move from left to right.

3. Press the ON/OFF button to turn the meter off. Rinse the electrode

with distilled water. Replace cap.

Storing Readings
1. After the reading is made press the MODE/HOLD button to store

the current reading. The

HOLD

icon will be displayed. The storage

location number will also be displayed followed by the reading
being stored.

2. Press the MODE/HOLD

button to return to normal
operation.

3. Repeat step 1 to store the next

reading.

4. If an attempt is made to store

more than 15 readings, the
stored readings (starting with
the first reading) will be overwritten.

Recalling Stored Readings
NOTE: First ensure that the

HOLD

symbol is not displayed. If it is, exit

the HOLD function by pressing the MODE/HOLD button.
1. Press the CAL/RECALL button once and then press the MODE

button immediately after

CAL

is displayed; the location number (1

through 15) will flash. If the CAL mode is accidentally accessed
(display flashing), press the CAL/RECALL button again to exit.

2. The last stored reading taken will be displayed first. To advance

through the stored readings, press the MODE/HOLD button. The
location number is displayed first, followed by the reading stored in
that location.

3. To exit the recall mode, press the CAL/RECALL button and the

TRACER will return to normal operation.

If the batteries are removed, any stored readings will be discarded.

7

Advertising