LaMotte 5 Series TDS Meter User Manual

Page 15

Advertising
background image

Preparing the Meter for Calibration
Before starting calibration, be sure the meter is in the correct measurement
mode.
For best results, select a standard value close to the value of the sample that is
being measured. Alternatively use a calibration solution value that is
approximately 2/3 the full-scale value of the measurement range that is being
used. For example, in the 0 to 2000 µS conductivity range, use a 1413 µS
solution for calibration.
Calibrate all measurement ranges to ensure the highest accuracy throughout
the entire measurement range. Note that the CON 6/TDS 6 meter will not
accept calibration values less than 40 µS/cm (20 ppm). All new calibration
values will automatically override existing data.
If solutions with Conductivity lower than 100 µS/cm, or TDS lower than 50
ppm, are being measured the meter should be calibrated at least once a week
for optimum accuracy. If measurements are in the mid-ranges, and the probe is
washed with deionized water and stored dry, the meter can be calibrated once a
month. If measurements are taken at extreme temperatures, the meter should
be calibrated at least once a week.
Fresh conductivity standard solutions should be used during calibration. Do
not reuse standard solutions. They could be contaminated and affect the
calibration and accuracy of the measurements. Store solutions in a dry and cool
environment if possible.
Always rinse the probe with either deionized water or rinse solution before and
after each calibration/sample measurement to avoid cross-contamination. For
details refer to Probe Care and Maintenance.

NOTE: These meters are factory set to a temperature coefficient of 2.1% per
°C. For most applications this will provide good results.
To set the temperature coefficient to different value, see Temperature
Coefficient.
Also, see Appendix 3 - Calculating the Temperature Coefficient to determine
the appropriate temperature coefficient for any solution.

NOTE: The factory default value for normalization temperature is 25 °C. If a
value other than 25 °C is needed, see Normalization Temperature.