LaMotte CDS 5000 Conductivity Meter User Manual

Page 8

Advertising
background image

TESTING WATER
CONDUCTIVITY
1. Press “°C” button to turn the meter on.
2. Set TDS and SLOPE knobs to the 12 o’clock position.

NOTE: If meter was calibrated according to procedure on page 7,

leave SLOPE knob set in the same position.

3. Insert temperature and conductivity probes at least 1/2" into sample.
4. Gently stir with conductivity probe until reading stabilizes. Press

mS/ppm” button. Record reading as msiemens/cm.

5. If a 1 appears on the far left side of the display, the reading is out of

range. Repeat procedure using “mS/ppt” button. Record as
msiemens/cm. To convert to

msiemens/cm, multiply reading by 1000.

Record as

msiemens/cm.

6. Press “OFF” button when finished testing. Rinse probe with distilled

water and dry thoroughly before storing.

TOTAL DISSOLVED SOLIDS
1. Set TDS knob to desired multiplication factor.

NOTE: 0.7 is a commonly used multiplication factor.

2. Press “°C” button to turn the meter on.
3. Set SLOPE knob to the 12 o’clock position.

NOTE: If meter was calibrated according to procedure on page 7,

leave SLOPE knob set in the same position.

4. Insert temperature and conductivity probes at least 1/2" into sample.
5. Gently stir with conductivity probe until reading stabilizes. Press

mS/ppm” button. Record reading as msiemens/cm.

6. If a 1 appears on the far left side of the display, the reading is out of

range. Repeat procedure using “mS/ppt” button. Record as
msiemens/cm. To convert to

msiemens/cm, multiply reading by 1000.

Record as

msiemens/cm.

7. Press “OFF” button when finished testing. Rinse probe with distilled

water and dry thoroughly before storing.

8