Electrode Replacement

Replace the electrode module when the Testr fails to calibrate, gives fluctuating
readings in buffers, shows error messages “E2" or ”OR", and the procedures in the
maintenance section fail to resolve the problem.

7

1.

2.

3.

4.

5.

Place the rubber gripper band

over the collar of the Testr.

Firmly grip the rubber gripper

band and the ribbed collar of

the Testr with dry hands.

Slowly twist counter-clockwise to

loosen the collar.

Unscrew the ribbed electrode

module collar and remove it. Save

the collar and the O-ring.

Pull the electrode module

straight out from the bottom

of the Testr.

Align the four tabs on the new

electrode module to match the

four slots on the Testr.

ON

OFF

CAL

HOLD

CON

WATERPROOF

pHTestr 1

ON

OFF

CAL

HOLD

CON

WATERPROOF

pHTestr 1

ON

OFF

CAL

HOLD

CON

WATERPROOF

pHTestr 1

ON

OFF

CAL

HOLD

CON

WATERPROOF

pHTestr 1

ON

OFF

CAL

HOLD

CON

WATERPROOF

pHTestr 1

ON

OFF

CAL

HOLD

CON

WATERPROOF

pHTestr 1

6.

small o-ring (hidden)