LaMotte TDS 5 Meter User Manual

Page 7

Advertising
background image

18

TDS 5 METER

TDS 5 METER

7

Meter Specifications

Range:

0.0 to 99.9 ppm

0.0 to 50°C

100 to 999 ppm
1.00 to 9.99 ppt
Auto Ranging

Resolution:

0.1 ppm

0.1°C

1 ppm
0.01 ppt

Accuracy:

+/- 2 ppm

+/- 0.5°C

+/- 20 ppm
+/- 0.2 ppt
(+/- 2% of full scale)

Calibration:

Up to 3 points

offset 0.1°C increments

TDS Factor:

User adjustable 0.5 to 0.85

Cell Constant:

1

Auto Shutoff:

after 17 minutes

Automatic Temp.
Comp.:

0.0 to 50°C

Reference Temperature: Factory set at 25°C
Temperature Coefficient: Factory set at 2% per °C
Keys:

ON/OFF, HOLD/ENTER, MODE/INC, CAL,
s/t

Power:

4 AAA Alkaline Batteries

Accessories

Description

Order #

TDS/Temp Probe

5-0042

Conductivity Standard, 52 ppm TDS, 500 mL

6416-L

Conductivity Standard, 503 ppm TDS, 500 mL

6417-L

Conductivity Standard, 989 ppm TDS, 500 mL

6354-L

Conductivity Standard, 4668 ppm TDS, 500 mL

6418-L

Protective Rubber Boot

5-0040

Error Messages

Error
Message

Indicates

Solution

E 2

Microprocessor

Error

Recalibrate instrument

E 3

Microprocessor

Error

Turn meter off and then on
again. Contact LaMotte for
return authorization

Ur

Temperature under range Warm sample

Faulty sensor

Contact LaMotte for return
authorization

Or

TDS over range

Dilute sample

Temperature over range

Cool sample

Faulty temperature sensor Contact LaMotte for return

authorization

Advertising