LaMotte AQ-3 Freshwater Aquaculture Kit User Manual

Fresh water aquaculture test kit

Advertising
background image

FRESH WATER

AQUACULTURE TEST KIT

MODEL AQ-3 CODE 3634-03

NOTE:

It is important to read the instruction manual

before attempting to perform the tests with the short

form instructions provided below.

*WARNING:

Reagents marked with a * are considered

hazardous substances. Material Safety Data Sheets (MSDS)

are supplied for these reagents. For your safety, read label

and accompanying MSDS before using.

To order individual reagents or test kit components, use the

specified code number.

AMMONIA NITROGEN

1.

Fill test tube (0106) to 5 mL line with sample water.

2.

Add 4 drops Ammonia Nitrogen Reagent #1

(4797WT). Cap and mix. Wait 1 minute.

3.

Add 12 drops of *Ammonia Nitrogen Reagent #2

(4798WT). Cap and mix. Wait 5 minutes.

4.

Insert test tube into Octa-Slide Viewer (1100) with

Ammonia Nitrogen Octa-Slide Bar (3438) inserted.

5.

Record as ppm Ammonia Nitrogen (NH

3

N).

NITRITE NITROGEN

1.

Fill test tube (0106) to 2.5 mL line with sample water.

2.

Dilute to 5 mL line with *Mixed Acid Reagent

(V-6278).

3.

Use 0.1g spoon (0699) to add 0.1g of *Color

Developing Reagent (V-6281). Cap and mix for 1

minute. Wait 5 minutes.

4.

Insert test tube into Octa-Slide Viewer (1100) with

Nitrite Nitrogen Octa-Slide Bar (3437) inserted.

5.

Record as ppm Nitrite Nitrogen (NO

2

N).

ALKALINITY

1.

Fill titration tube (0726) to 5 mL line with sample

water.

2.

Add 4 drops of *BCG-MR Indicator (2311-EG). Cap

and mix. Sample will turn blue-green.

@# = 3.

Fill Direct Reading Titrator (0382) with *Alkalinity

Titration Reagent B (4493DR).

4.

Titrate sample until blue-green color changes to pink.

5.

Record as ppm Alkalinity (CaCO

3

).

CARBON DIOXIDE

1.

Fill titration tube (0726) to 20 mL line with sample

water.

2.

Add 2 drops *Phenolphthalein Indicator, 1% (2246). If

sample turns red, no free carbon dioxide is present. If

colorless, proceed to Step 3.

3.

Fill Direct Reading Titrator (0380) with *Carbon

Dioxide Reagent B (4253DR).

4.

Titrate sample until faint pink color persists for 30

seconds.

5.

Record as ppm Carbon Dioxide (CO

2

).

CHLORIDE

1.

Fill titration tube (0726) to 15 mL line with sample

water.

2.

Add 1 drop *Phenolphthalein Indicator, 1% (2246). If

sample is colorless, proceed to Step 3. If sample turns

pink, add *Sulfuric Acid, 0.5N (6090) one drop at a

time until pink color disappears.

3.

Add 3 drops *Chloride Reagent #1 (4504). Cap and

mix. Sample will turn yellow.

4.

Fill Direct Reading Titrator (0382) with *Chloride

Reagent #2 (4505DR).

5.

Titrate sample until yellow color first changes to orange

or orange-red.

6.

Record as ppm Chloride (Cl).

HARDNESS

1.

Fill titration tube (0726) to 12.9 mL line with sample

water.

2.

Add 5 drops of *Hardness Reagent #5 (4483). Cap and

mix.

3.

Add 5 drops of Hardness Reagent #6 (4485). Cap and

mix. Sample will turn red.

4.

Fill Direct Reading Titrator (0382) with Hardness

Reagent #7 (4487DR).

5.

Titrate sample until red color changes to clear blue.

6.

Record as ppm Total Hardness (CaCO

3

).

LaMOTTE COM

PANY

PO BOX 329 ¨ CHESTERTOWN ¨ MARY LAND ¨ 21620

8/98

Advertising