Lamotte com pany – LaMotte CHLORINE IN DRINKING WATER TEST KIT SL-MW 3312 User Manual

Page 2

Advertising
background image

EPA AC CEPTED PROCEDURE
This test is EPA Accepted for NIPDWR monitoring. For compliance
monitoring a Check Standard should be prepared.
CHECK STANDARD PREP A RA TION: 1 ppm Equiv a lent Solution
1.

Dissolve 891 mg of potassium permanganate in 1000 mL of distilled
water in a volumetric flask (1000 ppm equivalent solution).

2.

Dilute 1 mL of this solution to 1000 mL with distilled water in a
volumetric flask. This solution is equivalent to 1 ppm Free Available
Chlorine.

PRO CE DURE

FREE AVAILABLE CHLORINE
1.

Fill a test tube (0106) to the 5 mL line with the sample water.

2.

Add one *Chlorine DPD #1R Tablet (6999A). Cap tube and mix
until tablet disintegrates.

3.

Insert test tube into the top of the Octa-Slide Viewer. Slide the
Chlorine Octa-Slide Bar (3405) into the viewer. Match sample color
to a color standard.

4.

Record as ppm Free Available Chlorine. Retain this sample if Total
and Combined Chlorine are to be determined.

TOTAL RE SID UAL & COMBINED CHLORINE
5.

Add one *Chlorine DPD #3 Tablet (6905A) to the test sample from
Step 4 above. Cap tube and mix until tablet dissolves.

6.

Insert test tube into the Octa-Slide Viewer. Match sample color to a
color standard.

7.

Record as ppm Total Residual Chlorine.

8.

To obtain reading in ppm Combined Chlorine, subtract Free
Available Chlorine from Total Residual Chlorine.

Combined Chlorine, ppm =

Total Residual Chlorine

_

Free Available Chlorine

LaMOTTE COM PANY

Helping Peo ple Solve An a lyt i cal Challenges

®

PO Box 329 • Chestertown • Mary land • 21620 • USA

800-344-3100 • 410-778-3100 (Out side U.S.A.) • Fax 410-778-6394

Visit us on the web at www.lamotte.com

1.11