PROCEDURE

FREE AVAILABLE CHLORINE

1. Fill a test tube (0106) to 5 mL line with sample water.

2. Add one *Chlorine DPD #1 Tablet (6999A). Cap and mix until tablet disintegrates.

Solution will turn pink if chlorine is present.

3. Insert test tube into the Octa-Slide 2 Viewer (1101). Slide the 0.2 to 3.0 ppm

Chlorine Octa-Slide 2 Bar (3401-01) into the viewer. Match sample color to a
color standard. Record as Reading A. Save sample for further tests.

Free Available Chlorine (ppm) = Reading A

MONOCHLORAMINE

4. Add one Chlorine DPD #2 Tablet (6904A) to sample from Step 3. Cap and mix

until tablet disintegrates.

5. Insert test tube into the Octa-Slide 2 Viewer (1101). Match sample color to a color

standard. Record as Reading B.

Monochloramine (ppm Chlorine) = Reading B - Reading A

DICHLORAMINE & TOTAL RESIDUAL CHLORINE

6. Add one *Chlorine DPD #3 Tablet (6905A) to the sample from Step 5. Cap and

mix until tablet disintegrates.

7. Insert test tube into the Octa-Slide 2 Viewer (1101). Match sample color to a color

standard. Record as Reading C.

Dichloramine (ppm Chlorine) = Reading C - Reading B

Total Residual Chlorine, ppm = Reading C

TOTAL RESIDUAL CHLORINE & COMBINED CHLORINE

1. Rinse test tube (0106) with sample water. Fill to 5 mL line with sample water.

2. Add one *Chlorine DPD #1 Tablet (6999A). Cap and mix until tablet

disintegrates.

3. Immediately insert test tube into the Octa-Slide Viewer (1101). Match sample

color to a color standard. Record as ppm Free Available Chlorine.

4. Add one *Chlorine DPD #3 Tablet (6905A) to the same sample. Cap and mix

until disintegrated.

5. Insert test tube into Octa-Slide 2 Viewer (1101). Match sample color to a color

standard. Record as ppm Total Residual Chlorine.

Combined Chlorine, ppm =

Total Residual Chlorine - Free Available Chlorine

TOTAL RESIDUAL CHLORINE

The *Chlorine DPD #4 Tablet determines Total Residual Chlorine and is used where a
differentiation in the type of chlorine is not required.

1. Rinse the test tube (0106) with sample water. Fill to 5 mL line with sample water.

2. Add one *Chlorine DPD #4 Tablet (6906A). Cap and mix until tablet

disintegrates.

3. Insert test tube into Octa-Slide 2 Viewer (1101). Match sample color to a color

standard. Record as ppm Total Residual Chlorine.