TO TAL RE SID UAL CHLO RINE & COM BINED CHLO RINE
1.

Rinse test tube (0106) with sample water. Fill to 5 mL line with
sample water.

2.

Add one *Chlorine DPD #1R Tablet (6999A). Cap and mix until
tablet disintegrates.

3.

Immediately insert test tube into the Octa-Slide Viewer (1100).
Match sample color to a color standard. Record as ppm Free
Available Chlorine.

4.

Add one *Chlorine DPD #3R Tablet (6905A) to the same sample.
Cap and mix until disintegrated.

5.

Insert test tube into Octa-Slide Viewer (1100). Match sample color
to a color standard. Record as ppm Total Residual Chlorine.

Combined Chlorine, ppm =

Total Residual Chlorine - Free Available Chlorine

TO TAL RE SID UAL CHLO RINE

The *Chlorine DPD #4R Tablet determines Total Residual Chlorine and
is used where a differentiation in the type of chlorine is not required.
1.

Rinse the test tube (0106) with sample water. Fill to 5 mL line with
sample water.

2.

Add one *Chlorine DPD #4R Tablet (6906A). Cap and mix until
tablet disintegrates.

3.

Insert test tube into Octa-Slide Viewer (1100). Match sample color
to a color standard. Record as ppm Total Residual Chlorine.