Lamotte com pany – LaMotte PHOSPHORUS IN SOIL SL-P User Manual

Page 4

Advertising
background image

USE OF THE OCTA-SLIDE VIEWER

The Octa-Slide Viewer should be held so
non-direct light enters through the back of the
comparator. With sample tube inserted at top,
slide the color bar through the viewer and match
with color standards.

PRO CE DURE
NOTE:

It is suggested for best possible results to carry a reagent blank

through the procedure to compensate for any color which may develop
within the reagents. The reading from the reagent blank is subtracted
from the reading of the test sample.
1.

Use the 1 mL pipet (0354) to add 1 mL of *Acid Extracting
Solution (6361) to a test tube (0701). Add deionized water to the
15 mL line.

2.

Use the 1.0 g spoon (0697) to add 3 measures of soil.

3.

Use the 0.5 mL pipet (0353) to add 0.5 mL of Charcoal
Suspension (5638). Cap vial and shake for 5 minutes.

4.

Use the funnel (0459) and filter paper (0465) to filter solution.
Collect the soil filtrate from both tubes in one vial (0989).

5.

Use the 1.0 mL pipet (0354) to add 1.0 mL of the soil filtrate to a
test tube (0106).

6.

Dilute the filtrate to the 5 mL line with deionized water.

7.

Use the 0.5 mL pipet (0353) to add 0.5 mL of *VM Phosphate
Reagent (4410). Cap tube and mix by inverting the tube several
times.

8.

Wait for 5 minutes.

9.

Use the plain pipet (0352) to add 2 drops of *Reducing Reagent
(6405). Cap and mix. Solution should turn blue in 10 seconds.

10.

Insert test tube into the top of the Octa-Slide Viewer (1100). Slide
the Phosphorus Octa-Slide (3423) into the Viewer. Match sample
color to a color standard. Record as lb/acre Phosphorus.

LaMOTTE COM PANY

Helping Peo ple Solve An a lyt i cal Chal lenges

SM

PO Box 329 • Chestertown • Mary land • 21620 • USA

800-344-3100 • 410-778-3100 (Out side USA) • Fax 410-778-6394

Visit us on the web at www.lamotte.com

12/01