LaMotte ALKALINITY TEST KIT DR-A User Manual

Al ka lin ity test kit

Advertising
background image

AL KA LIN ITY TEST KIT

MODEL DR-A • CODE 3467

QUAN TITY

CON TENTS

CODE

15 mL

*Alkalinity Indicator #1, pH 8.3

*3870-E

15 mL

Alkalinity Indicator #2, pH 4.5

3869-E

60 mL

*Alkalinity Titration Reagent B

*4493DR-H

1

Test Tube, 5-10-15 mL, glass, w/cap

0778

1

Direct Reading Titrator, 0-200 Range

0382

*WARNING:

Reagents marked with an * are considered to be potential

health hazards. To view or print a Material Safety Data Sheet (MSDS) for
these reagents go to www.lamotte.com. To obtain a printed copy, contact
LaMotte by e-mail, phone or fax.
To order a complete set of refill reagents, order as R-3467. To order individual
reagents or test kit components, use the specified code number.

NOTE:

Carefully read the instruction manual for the LaMotte Direct

Reading Titrator before performing the titration described below. The
Titrator is calibrated in terms of alkalinity expressed as parts per million
(ppm) calcium carbonate (CaCO

3

). Each minor division on the Titrator

scale equals 4 ppm CaCO

3

.

EPA AC CEPTED PRO CE DURE
To qualify as an EPA accepted test and to achieve greatest accuracy, the
normality of the *Alkalinity Reagent B should be checked periodically.
Standardize *Alkalinity Reagent B by titrating against 0.05N Na

2

CO

3

;

1 mL = 1.00 mg CaCO

3

. Should a variance of greater than 0.3 mL of

*Alkalinity Reagent B occur, discard and order a fresh bottle of
*Alkalinity Reagent B (Code 4493DR-H). A CaCO

3

equivalent

standard of 100 ppm is also available (Code 6199-H). Refer to Standard
Methods

, 16th edition, 403, 1985.

EPA