LaMotte BROMINE/CHLORINE TEST KIT CL-BR User Manual

Bro mine/chlo rine test kit

Advertising
background image

BRO MINE/CHLO RINE TEST KIT

DIRECT READING TITRATOR METHOD

MODEL CL-BR • CODE 3624

QUANTITY

CONTENTS

CODE

5 g

*DPD #1 Powder

*6807-C

50

*Chlorine DPD #3R Tablets

*6905A-H

60 mL

Chlorine/Bromine Titrant

3992DR-H

1

Test Tube, 10 & 22.5 mL, w/cap

0753

1

Direct Reading Titrator, 0-10 Range

0377

1

Spoon, 0.1 g, plastic

0699

1

Pipet, 1.0 mL, plastic

0354

*WARNING: Reagents marked with an * are considered to be potential health
hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents go to
www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specified code number.

Read the Direct Reading Titrator Manual before proceeding. Each minor
division equals 0.2 ppm.

PRO CE DURE
BROMINE (0 - 10 ppm)
1. Fill the test tube (0753) to the 22.5 mL line with sample water.
2. Use the 0.1 g spoon (0699) to add 0.1 g of *DPD #1 Powder (6807).

Cap and gently swirl until powder dissolves. Solution will turn red if
bromine is present.

3. Fill the Direct Reading Titrator (0377) with the Chlorine/Bromine

Titrant (3992). Insert the Titrator tip into the center hole of the test
tube cap.

4. While gently swirling the tube, slowly press plunger to titrate until red

color completely disappears.

5. Read the test result directly from the scale where the large ring on the

Titrator meets the Titrator barrel. Record as ppm Bromine.

63624-5/10