PROCEDURE

NOTE: If sample water is cloudy, fi lter before testing.

1.0 - 10.0 ppm PHOSPHATE (PO

4

)

1. Fill a test tube (0106) to the 5 mL line with sample water.
2. Use the 1.0 mL pipet (0354) to add 1.0 mL of *VM Phosphate Reagent (4410).

Stopper and mix by inverting several times. Wait 5 minutes.

3. Use the plain pipet (0352) to add 3 drops of *Reducing Reagent (6405). Cap and

mix. Color will develop in 10 seconds.

4. Insert the Phosphate Octa-Slide 2 Bar (4414-01) into the Octa-Slide 2 Viewer

(1101).

5. Insert test tube into the Octa-Slide 2 Viewer. Match sample color to a color stan-

dard. Record as ppm Phosphate (PO

4

).

10 - 100 ppm PHOSPHATE (PO

4

): 1 to 10 DILUTION

1. Use the 0.5 mL pipet (0353) to add 0.5 mL of sample water to a test tube (0106).
2. Dilute to 5 mL line with phosphate-free (distilled) water.
3. Use the 1.0 mL pipet to add 1.0 mL of *VM Phosphate Reagent (4410). Cap and

mix by inverting several times. Wait 5 minutes.

4. Use the plain pipet (0352) to add 3 drops of *Reducing Reagent. Cap and mix.

Color will develop within 10 seconds.

5. Insert the Phosphate Octa-Slide 2 Bar (4414-01) into the Octa-Slide 2 Viewer

(1101)

6. Insert test tube into the Octa-Slide 2 Viewer. Match sample color to a color stan-

dard. Multiply by 10. Record as ppm Phosphate (PO

4

).

CAUTION: Do not use soaps or detergents containing phosphates to wash test tubes.