Total hardness test procedure, Caco, Gpg caco = ppm caco x 0.058 – LaMotte HARDNESS TEST KIT 4824-DR-LT User Manual

Page 2: While gently swirling the

Advertising
background image

TOTAL HARDNESS TEST PROCEDURE

If the plunger reaches the bottom line on the scale

(200 ppm) before the endpoint color change occurs, refill

Read the test

result directly

from the scale

where the large

ring on the

Titrator meets the

Titrator barrel.

Record as ppm

Total Hardness as

(CaCO ).

3

7.

NOTE:

the Titrator and continue the titration. When recording the test result,

be sure to include the value of the original amount of reagent dispensed

(200 ppm).
Parts per million CaCO may be converted to grains per gallon (gpg)

3

CaCO :

3

gpg CaCO = ppm CaCO x 0.058

3

3

EXAMPLE:
Plunger tip is

2 minor divisions

below line 80.
Test result is:

80 + (2 divisions x 4)

= 88 ppm.

80

100

Result:

88 ppm

120

2.

5.

Fill

Direct Reading

Titrator (0382)

with Hardness

Reagent #7

(4487DR).

Insert the

Titrator into

the center

hole of the

test tube cap.

1.

3.

4.

tube, slowly press

the plunger to

titrate until the

red color changes

to clear blue.

Add 5 drops

of *Hardness

Reagent #5

(4483).

Swirl to mix.

Add one Hardness

Reagent #6 Tablet

(4484). Cap and swirl

Fill a test

tube (0608)

to the 12.9

mL line with

the sample

water.

until tablet disintegrates.

Solution will turn red if

hardness is present.

If solution is blue,

there is no

measurable

amount of

hardness.

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

While gently

swirling the

6.

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

Advertising