EPA AC CEPTED PROCEDURE
EPA Accepted for NPDWR compliance monitoring.

For compliance

monitoring a Check Standard should be prepared. A fresh solution
should be prepared daily.
CHECK STANDARD PREP A RA TION: 1 ppm equiv a lent so lu tion
1.

Place 891 mg potassium permanganate in a 1 liter volumetric flask.
Dilute to 1 liter with distilled or deionized water. This is the stock
solution.
NOTE:

This stock solution should be prepared fresh daily.

2.

Transfer 1 mL to a 1 liter volumetric flask. Dilute to 1 liter with
distilled or deionized water. This solution is equivalent to 1 ppm
Free Available Chlorine.

PRO CE DURE

FREE AVAIL ABLE CHLO RINE
1.

Rinse a test tube with sample water. Fill to 5 mL line with the
sample water.

2.

Add one *Chlorine DPD #1R Tablet (6999A). Cap and shake until
tablet disintegrates.

3.

Immediately insert test tube into Chlorine DPD Comparator
(6902). Match sample color to a color standard. Record as ppm
Free Available Chlorine.
NOTE:

Save sample for total residual and combined chlorine tests.

TO TAL RE SID UAL CHLO RINE & COM BINED CHLO RINE
4.

Add one *Chlorine DPD #3R Tablet (6905A) to the sample from
Step 3. Cap and shake until tablet disintegrates.

5.

Insert test tube into Chlorine DPD Comparator (6902). Match
sample color to a color standard. Record as ppm Total Residual
Chlorine.

6.

Subtract Free Available Chlorine from Total Residual Chlorine.
Record as ppm Combined Chlorine.

Combined Chlorine, ppm =

Total Residual Chlorine - Free Available Chlorine

NOTE:

Thoroughly clean and rinse test tubes after each test.

LaMOTTE COM PANY

Helping People Solve Analytical Challenges

®

PO Box 329 • Chestertown • Mary land • 21620 • USA

800-344-3100 • 410-778-3100 (Out side U.S.A.) • Fax 410-778-6394

Visit us on the web at www.lamotte.com

66817-4.11