Lamotte com pany – LaMotte PHOSPHATE TEST KIT P-POR User Manual

Page 2

Advertising
background image

PRO CE DURE

HIGH RANGE: 10-100 PPM PHOS PHATE (PO

4

)

1:10 DI LU TION
1.

Use the 1.0 mL pipet (0354) to add 1.0 mL of sample water to the
test tube (0970-S).

2.

Dilute to 10 mL line with distilled or phosphate free water.

3.

Use the 0.5 mL pipet (0353) to carefully add 0.5 mL of *Phosphate
Reagent A (4606).

4.

Add one *Phosphate B/C Tablet (4607). Cap and gently mix until
the tablet is dissolved. Wait 4 minutes.

5.

Fill a 5 mL test tube (0230) to the 5 mL line with this reacted
sample.

6.

Insert test tube into Phosphate Comparator (7899). Match sample
color to a color standard. Record as ppm Phosphate.
NOTE:

If sample color is lighter than the lightest color standard,

use the following procedure.

LOW RANGE: 1-10 PPM PHOS PHATE (PO

4

)

UN DI LUTED SAM PLE
1.

Fill the test tube (0970-S) to the 10 mL line with sample water.

2.

Use the 0.5 mL pipet (0353) to carefully add 0.5 mL of *Phosphate
Reagent A (4606).

3.

Add one *Phosphate B/C Tablet (4607). Cap and gently mix until
the tablet is dissolved. Wait 4 minutes.

4.

Fill a 5 mL test tube (0230) to the 5 mL line with this reacted
sample.

5.

Insert test tube into Phosphate Comparator (7899). Match sample
color to a color standard.
IMPORTANT:

Divide the value obtained from the comparator by

10. Record as ppm Phosphate.

CAUTION:

Do not use soaps or detergents containing phosphate to

wash test tubes.

LaMOTTE COM PANY

Helping Peo ple Solve An a lyt i cal Chal lenges

®

PO Box 329 • Chestertown • Mary land • 21620 • USA

800-344-3100 • 410-778-3100 (Out side U.S.A.) • Fax 410-778-6394

Visit us on the web at www.lamotte.com

7.09