PRO CE DURE

Concentration

Sample Size

DRT multiplication factor

0-10%

0.5 mL

1

0-5%

1.0 mL

0.5

0-1%

5 mL

0.1

0-0.5%
0-5,000 ppm

10 mL

0.05 for %
500 for ppm

0-0.25%
0-2,500 ppm

20 mL

0.025 for %
250 for ppm

0-0.2%
0-2,000 ppm

25 mL

0.020 for %
200 for ppm

1.

Add the sample to be tested to the test tube (0608). For a 0.5 mL or
1.0 mL sam ple size, fill the tube to 10 mL with tap wa ter and use the
0.5 mL pipet (0353) or 1.0 mL pipet (0354) to add the appropriate
amount of sample. For sample sizes greater than 1.0 mL, fill the test
tube (0608) to the specified line.

2.

Use the 1.0 mL pipet with red tape (NA-0354) to add 1.0 mL of
*Hypochlorite Reagent A (7939). Swirl to mix.

3.

Use 0.1g spoon (0699) to add 0.1g *Potassium Iodide Crystals
(6809). Swirl to mix.

4.

Fill the Direct Reading Titrator (0377) with Hypochlorite Reagent D
(2790). See Direct Reading Titrator Manual (1649). Insert the
Titrator tip into the center hole of the test tube cap.

5.

While gently swirl ing the tube, slowly de press the plunger to ti trate
the sam ple un til the so lu tion changes from brown to col or less and
re mains col or less for 1 min ute. Read the test re sult di rectly from the
scale where the large ring on the Titrator meets the Titrator bar rel.
Mul ti ply titrator read ing by the ap pro pri ate fac tor listed in the ta ble
above. NOTE: The so lu tion will go through an in ter me di ate yel low
stage.

LaMOTTE COMPANY

Helping Peo ple Solve An a lyt i cal Chal lenges

®

PO Box 329 • Chestertown • Mary land • 21620 • USA

800-344-3100 • 410-778-3100 (Out side U.S.A.) • Fax 410-778-6394

Visit us on the web at www.lamotte.com

10.07