LaMotte MAGNESIUM HARDNESS TEST KIT AB-HCM User Manual

LaMotte Equipment

Advertising
background image

TO TAL, CAL CIUM AND

MAG NE SIUM HARD NESS TEST KIT

AUTOMATIC BURET METHOD

MODEL AB-HCM • CODE 7173-01

QUANTITY

CONTENTS

CODE

100

Hardness Reagent #6 Tablets

4484-J

100 mL

*Sodium Hydroxide Reagent

w/Metal Inhibitors

*4259-J

100

Calcium Hardness Indicator Tablets

T-5250-J

2 x 250 mL

Hardness Titration Reagent

4257-K

30 mL

*Hardness Reagent #5

*4483-G

1

Erlenmeyer Flask, 125 mL, glass

0431

1

Graduated Cylinder, 25 mL, glass

0417

1

Pipet, 1.0 mL, plastic

0354

1

Automatic Buret

0427

*WARNING:

Reagents marked with an * are considered to be potential health

hazards. To view or print a Material Safety Data Sheet (MSDS) for these
reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by
e-mail, phone or fax.
To order individual reagents of test kit components, use the specified code
number.

Read the Automatic Buret Instruction Manual before proceeding.

PRO CE DURE
TOTAL HARDNESS
1.

Fill graduated cylinder (0417) to the 25 mL line with sample water.
Transfer to the clean 125 mL Erlenmeyer flask (0431).

2.

Add 10 drops of *Hardness Reagent #5 (4483). Mix.

3.

Add one Hardness Reagent #6 Tablet (4484). Mix until tablet
disintegrates. Sample should turn red.

Advertising