PRO CE DURE

1 drop = 0.1%
1.

Fill test tube (0715) to 5 mL line with sample.

2.

Add one drop of *Phenolphthalein Indicator, 0.5% (2258). Mix.

3.

While gently swirling tube, add *Sodium Hydroxide, 2.4N (5648)
one drop at a time until solution turns pink. Count the number of
drops added. Hold bottle vertically.

4.

Multiply the number of drops used in Step 3 by 0.1. Record as %
Phosphoric Acid or Sulfuric Acid.
To receive result as % hydrochloric acid, multiply number of drops
used in Step 3 by 0.076. Record as % Hydrochloric Acid.

1 drop = 1%
1.

Use 0.5 mL pipet (0353) to add 0.5 mL of sample to test tube (0715).

2.

Dilute to 5 mL line with tap water.

3.

Add one drop of *Phenolphthalein Indicator, 0.5% (2258). Mix.

4.

While gently swirling tube, add *Sodium Hydroxide, 2.4N (5648)
one drop at a time until solution turns pink. Count the number of
drops added. Hold bottle vertically.

5.

Multiply the number of drops used in Step 4 by 1. Record as %
Phosphoric Acid or Sulfuric Acid.
To receive result as % hydrochloric acid, multiply number of drops
used in Step 3 by 0.76. Record as % Hydrochloric Acid.

LaMOTTE COM PANY

Helping Peo ple Solve An a lyt i cal Challenges

®

PO Box 329 • Chestertown • Mary land • 21620 • USA

800-344-3100 • 410-778-3100 (Out side U.S.A.) • Fax 410-778-6394

Visit us on the web at www.lamotte.com

3.11