PROCEDURE

1. Use the 1.0 mL pipet (0354) to add 2 mL of sample water to a 5 mL test tube

(0106).

2. Use water from the Demineralizer Bottle (1151) to fi ll test tube to the 5 mL line.

Cap and mix.

3. Add 5 drops of *Zinc Conditioning Reagent (7361). Cap and mix. Wait one

minute. This reagent will eliminate copper interference and adjust the pH of the
sample water to pH 9. In water samples where the pH is very low, adjust the pH
of the sample to approximately 9 before adding the *Zinc Conditioning Reagent
(7361).

4. Use the 0.5g spoon (0698) to add one level measure of *Zinc Reagent Powder

(7393). Cap and shake for 15 seconds. Do not shake for longer than 15 seconds,
even if some powder remains undissolved. Wait one minute. Do not mix.

5. Immediately, insert the Zinc Octa-Slide 2 Bar (7392-01) into the Octa-Slide 2

Viewer (1101). If reading is darker than 10 ppm standard see procedure for High
Range Readings.

6. Insert the test tube into the Octa-Slide Viewer.

7. Match the sample color to a color standard. Read within 30 seconds. Record as

ppm Zinc.

NOTE: Slight variations in the amount of *Zinc Reagent Powder (7393) added
in Step 4 may cause the intensity, but not the shade, of the color reaction to vary.
Since each color standard exhibits a distinct shade of color, this should not hinder
the ability to get an accurate reading.

HIGHER RANGE READINGS (0-20.0 ppm)

If reading is darker than the 10 ppm color standard repeat test on a diluted sample.

1. Use the 1.0 mL pipet (0354) to add 1 mL of sample water to a test tube (0106).

2. Follow Steps 2 through 7 above. Multiply reading by 2. Record as ppm Zinc.

LOWER RANGE READINGS (0-4.0 ppm)

For more precise measurements between 0 and 4 ppm.

1. Fill test tube (0106) to 5 mL line with sample water.

2. Follow Steps 3 through 7 above. Multiply reading by 0.4. Record results as ppm

Zinc.

LaMOTTE COMPANY

Helping People Solve Analytical Challenges

®

PO Box 329 • Chestertown • Maryland • 21620 • USA

800-344-3100 • 410-778-3100 (Outside U.S.A.) • Fax 410-778-6394

Visit us on the web at www.lamotte.com

67391-02 01/12