LaMotte BOD5 TEST KIT BOD User Manual

Page 3

Advertising
background image

TABLE I: EXPECTED BOD

5

RANGE

Sample added to 300 mL

bottle (mL)

Minimum

(mg/L)

Maximum

(mg/L)

3

210

560

6

105

280

9

70

187

12

53

140

15

42

112

18

35

94

21

30

80

24

26

70

27

24

62

30

21

56

45

14

37

60

11

28

75

8

22

150

4

12

2.

Fill two 300 mL BOD bottles (2-2033) about half full with dilution water.
With an appropriately sized pipet or graduated cylinder dispense the
precalculated amount of the sample into each of the two 300 mL BOD
bottles. Fill each bottle with dilution water and insert stopper. Exclude all air
bubbles.

3.

Fill an additional two 300 mL BOD bottles (2-2033) only with dilution
water. Insert stoppers.

4.

Incubate one bottle containing diluted sample (Step 2) and one bottle
containing only dilution water (Step 3) at 20°C for 5 days. Ensure that the
area above the stopper contains distilled water. Check daily or use bottle
covers to prevent evaporation (not supplied).

5.

Run a dissolved oxygen determination on the remaining BOD bottle from
Step 2. Record the initial dissolved oxygen content as D

1

. Run another

dissolved oxygen determination on the remaining BOD bottle from Step 3.
Record initial dissolved oxygen content for control as C

1

. See Dissolved

Oxygen Determination Procedure.