PRO CE DURE
1.

Pour 10 mL of *Ammonia Nitrogen Absorbing Solution (7737)
into impinging apparatus. Connect impinging apparatus to intake of
air sampling pump.
NOTE:

Make sure the long tube is immersed in the absorbing

solution.

2.

Set flowmeter to sample at a rate of 1.0 Lpm for 10 minutes or until
a measurable amount of ammonia is absorbed.

3.

At the end of the sampling period, add contents of impinger to the
5 mL line of test tube (0230).

4.

Add 2 drops of Ammonia Nitrogen Reagent #1 (4797WT). Cap
and mix.

5.

Add 8 drops of *Ammonia Nitrogen Reagent #2 (4798WT). Cap
and mix. Solution will turn yellow to brown color if ammonia is
present.

6.

Place test tube into Ammonia In Air Comparator (7736). Match
sample color to index of color standards. Note the index number
which gives the proper color match. Record results as ppm
Ammonia Nitrogen.
NOTE:

To convert ammonia nitrogen to ammonia, multiply chart

value by 1.2. Record as ppm Ammonia.

AMMONIA IN AIR CALIBRATION CHART**

TIME

COMPARATOR INDEX NUMBER

(min)

1

2

3

4

5

6

7

8

5

3.43

6.86

10.30

13.73

17.16

20.59

24.02

27.46

10

1.72

3.43

5.15

6.86

8.58

10.30

12.01

13.73

15

1.15

2.28

3.43

4.58

5.71

6.86

8.02

9.14

20

0.86

1.72

2.58

3.43

4.30

5.15

6.01

6.86

**Values in ppm