5 charts/pens – Monarch Instrument DC 2000 User Manual

Page 72

Advertising
background image

Page 5-6

Chapter 5 Programming

Level”. Use the UP (

↑) or DOWN (↓) Arrow Keys to select a Dim Level and press ENTER. The Dim Level

selection will be one level less than the Brightness Level you set. For example, if you select “2” as the Bright
Level, you will be given the choice of 1 or OFF for the Dim Level. Next, enter the time out interval in minutes -
“minutes

→XXX←” and press ENTER. Setting the time out value to 0 never dims the backlight.

The display will go to the “Bright Level” when ACK button is flashing or when the touch screen is pressed. If an
alarm regularly goes off within the screen dimmer time out interval, the display will never go to the “Dim Level”.

NOTE: When selecting “OFF” as the Dim Level, the display does not turn off when programming.

5.5 Charts/Pens

This Programming menu item allows the user to program parameters directly affecting charts or pens. The Chart/
Pens programming menu is shown in Figure 5-4. Each programming item shown has further programming options
as listed below.

Speed ………

Set speeds
Autospeed

Scales ………

Scale #
Scale type
Scale ends
Scale grid
Scale units

Pens ………..

Pens assign
Abnorm. pen
Trace Width

Direction ……

Vertical
Horizontal

5.5.1 Speed

This menu item establishes the Virtual Chart Speed and Units (Inches/Hour - Standard or Millimeters/Hour -
Metric) at which the virtual chart paper will advance. Display Update Rate or Virtual Chart Speed may be set
to Autospeed which enables chart speed to be changed via Alarm condition or external input. Set Speeds
enables the user to select the virtual chart speed. Virtual Chart Speeds are programmable within the following
limits:

Standard (0.5 in/hr to 600 in/hr)

Metric (10 mm/hr to 15,000 mm/hr)

increments of 0.5 from 0.5-10

increments of 5 from 10-100

increments of 5 from 10-100

increments of 50 from 100-1000

increments of 50 from 100-600

increments of 500 from 1000-10,000

increments of 1000 from 10,000-15,000

Programming the Display Update Rates or virtual chart speeds - When in the Chart/Pens programming
menu, use the UP (

↑) or DOWN (↓) Arrow Keys to highlight “Speed” and press ENTER to select then use the

UP (

↑) or DOWN (↓) Arrow Keys to highlight “Set speeds”, or “Autospeed”. Press ENTER to accept the

displayed speed control of Autospeed or Set speeds. If Autospeed is selected, use the YES or NO keys to
change the display to read “autospeed = NO” or “autospeed = YES”. When the desired Autospeed function is
displayed, press ENTER to invoke the autospeed selection and return to the “Speed” menu item.

If Set speeds is selected, either the “standard*” or “metric*” menu item will appear. Use the UP (

↑) or DOWN

(

↓) Arrow Keys to select the desired selection. Press ENTER at the desired speed type to set the low chart

speed. When “lo spd = X.X” appears, use the UP (

↑) or DOWN (↓) Arrow Keys to select the desired low chart

speed. Press ENTER and “hi spd= X.X” will be displayed. Use the UP (

↑) or DOWN (↓) Arrow Keys to select

the desired high chart speed. Press ENTER when programming high chart speed is complete and the display
returns to the “Set speeds” menu item. Press EXIT three times to return to the Main Programming Button Bar.

Scales
Pens
Direction

Speed

Figure 5-4

Chart/Pens Menu

Advertising