Installation – Amico Majestic Series Double Tier Recessed Headwall User Manual
Page 4

4
Amico Corporation
Installation
1. Connect on-site medical gas and electrical sources to the unit as per local building codes. The Headwall is
manifolded for a single point of connection for each medical gas. Electrical devices are terminated at junction
locations at the top of the unit. See Figure 6 for reference. See project specific rough in drawings for exact rough-
in locations.
2. Proceed with drywall installation. See Figure 7 for drywall setup.
3. Remove the fascias from the frame using the lift method (See Figure 8).
Figure 6
medical gas
from source
medical gas
terminations
from unit
electrical conduits
from source
connectors
provided with the
console assembly
Figure 7
Front frame shown
for reference only. It
will get installed after
drywall and paint.
Drywall
Drywall
Figure 8