Installation caution – Amico LDR Headwall User Manual
Page 3

Installation
Caution
Secure the LDR metal frame to the wall structure. Use hardware (by others) compliant to
applicable building codes. Holes in mounting frame are to be drilled according to wall structure
(stud) location. Back plate is also recommended.
Make all rough-in connections between wall and LDR frame for piping and electrical services.
Electrical services are terminated at junction boxes, and gas terminations are located according
to rough-in drawings (provided) to facilitate installation.
Please see details A and B for reference.
Do not rest the cabinet on its movable bottom tray. Always lay cabinet on its back to avoid
damage.
1.
2.
www.amico.com
2
med gas termination
(see detail A)
frame assembly
electrical termination box
(see detail B)
studs
drywall