EMC-Inspection Report

Guidance and Manufacturer’s Declaration – Electromagnetic Emissions

The Gamma 50 LED is intended for use in the electromagnetic environment specified below. The customer or the user of the

Gamma 50 LED should assure that it is used in such an environment.

Emission test

Compliance

Electromagnetic environment - guidance

RF emissions

CISPR 11

Group 1

The Gamma 50 LED uses RF energy only for its internal

function. Therefore, its RF emissions are very low and are

not likely to cause any interference in nearby electronic

equipment.

RF emissions

CISPR 11

Class B

The Gamma 50 LED is suitable for use in all establishments,

including domestic establishments and those directly

connected to the public low-voltage power supply network

that supplies buildings used for domestic purposes.

Harmonic emissions

IEC 61000-3-2

Class C

Voltage fluctuations / flicker emissions

IEC 61000-3-3

Complies

Guidance and Manufacturer’s Declaration – Electromagnetic Immunity

The Gamma 50 LED is intended for use in the electromagnetic environment specified below. The customer or the user of Gamma

50 LED should assure that it is used in such an environment.

Immunity test standard

IEC 60601 test level

Compliance level

Electromagnetic environment - guidance

Electrostatic discharge

(ESD)

IEC 61000-4-2

±6 kV contact
±8 kV contact

±8 kV air

±15 kV air

Floors should be wood, concrete or ceramic tile.

If floors are covered with synthetic material, the

relative humidity should be at least 30%

Electrical fast transient /

burst

IEC 61000-4-4

±2 kV for power

supply lines

±2 kV for power

supply lines

Mains power quality should be that of a typical

commercial or hospital environment.

Surge

IEC 61000-4-5

±1 kV line to line

±2 kV line to earth

±1 kV line to line

±2 kV line to earth

Mains power quality should be that of a typical

commercial or hospital environment.

Table 1: Emission

Table 2: Immunity (All Devices)