Agss flow control adjustment, British standard a.g.s.s – Amico British Standard Outlet Slim User Manual

Page 11

Advertising
background image

www.amico.com

11

AGSS Flow Control Adjustment

Surface Mount

Box

Flow Control

Screw

(3mm Allen Key)

Valve

A.G.S.S.

Threaded Nose

www.amico.com

British Standard A.G.S.S.

1. Insert a 3 mm Allen Key into the front of the Check Valve.

2. Adjust the Flow Control Screw to desired setting.

Referencing Part Numbers:

O-BSQSL-SUR-AGS
O-BSQSL-CON-AGS