Installation – Amico Dome Loaded Manifold CSA User Manual

Page 12

Advertising
background image

12

Amico Pipeline

Installation

PRECAUTIONS

• Tampering with gas specific connections shall be prohibited. Do not alter, remove or modify gas specific connections.

• Keep all manifold parts, tools and work surfaces free of oil, grease and dirt. These and other flammable materials

may ignite when exposed to high pressure oxygen or nitrous oxide.

• Do not use chemicals, lubricants or sealants unless specified in these instructions.

• Before connecting cylinder to manifold, momentarily open and close cylinder valve to blow out dirt and debris.

• After connecting cylinder to manifold, open cylinder valve S-L-O-W-L-Y to allow heat of compression to dissipate.

• Do not use flame or “sniff” tests for leaks.

• Do not apply heat to any part of the manifold system.

• Always secure high-pressure cylinders with racks, straps, or chains. Unrestrained cylinders may fall over and

damage or break off cylinder valve.

• Do not repeatedly bend, sharply bend, or twist copper pigtails as damage to tubing may occur.

• After the manifold wall bracket has been mounted, one person alone should not attempt to lift and hang

the manifold cabinet.

• Do not put the manifold into operation until verified by a qualified person as per local standards.